Radiation Lab Field Pro

Validation: the engine against the ISO 4037 tables

Field Pro · updated 28 October 2026

A reference-field calculator is only useful if you know how far to trust it. This page lists what the engine is compared against, the deviations it produces, the reasons we can identify for them, and what has not been checked.

How the checks run

The comparisons are code, not a one-off spreadsheet. The unit-test suite (153 tests) recomputes the first half-value layer of every ISO 4037-1 quality and fails if any deviation leaves its envelope, and a report script writes the same numbers into the accuracy specification from the engine itself rather than by hand. Browser tests on a 1440 px desktop and a Pixel 7 profile then open the built site and read the screen: with the default inputs the HVL1 readout must show 0.57 mm Cu and the auto-match must name N-80, and five sample readings must produce a Type A coverage factor of 2.776.

First half-value layer, all 50 qualities

Each quality is built from its tabulated tube voltage, inherent and added filtration and a 2.5 m air path, and its HVL1 is compared with the 2.5 m column of ISO 4037-1:2019.

Mean energyQualitiesMean |deviation|Largest deviationEnvelope
below 40 keV171.26%H-10, −3.91%≤ 4%
40–150 keV220.67%H-280, +1.45%≤ 2%
150 keV and above111.02%N-300, +1.69%≤ 2%

For N-80, the default field, the full comparison reads:

CharacteristicISO 4037-1EngineDeviation
HVL1 (mm Cu)0.5800.5753−0.80%
HVL2 (mm Cu)0.6230.6185−0.72%
Homogeneity coefficient0.93100.9302−0.08%
Mean energy (keV)65.264.96−0.37%

Radionuclide fields

Spectrum-averaged conversion coefficients at normal incidence, in Sv/Gy, against the values ISO 4037-3:2019 recommends for the two nuclide qualities:

NuclideQuantityISO 4037-3EngineDeviation
Cs-137Hp(10)1.2101.216+0.52%
Cs-137H*(10)1.2001.203+0.26%
Cs-137H′(0.07)1.2101.200−0.82%
Co-60Hp(10)1.1501.153+0.27%
Co-60H*(10)1.1601.159−0.10%
Co-60H′(0.07)1.1501.162+1.03%

The nuclide data agree with the ENSDF evaluations shown in the IAEA Live Chart of Nuclides: 30.08 a and 661.657 keV at 85.1% for Cs-137; 1925.28 d (5.2711 a), 1173.228 keV at 99.85% and 1332.492 keV at 99.98% for Co-60. Carried through to the air-kerma rate constant, the engine gives 0.0758 and 0.3054 µGy·m²·MBq⁻¹·h⁻¹; the test suite holds them within 4% of the literature values 0.0771 and 0.3085, and they sit 1.7% and 1.0% below.

Where the deviations come from

No factor is fitted to make the engine meet the table: the emission data come straight from SpekPy and the attenuation data from NIST, and the positive deviations at N-300 and H-280 are reported as they are.

What is not validated

Sources

  1. ISO 4037-1:2019, Part 1: Radiation characteristics and production methods, Tables 2–6 (HVL1, HVL2 and mean energy at 1.0 m and 2.5 m).
  2. ISO 4037-3:2019, Part 3: Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence.
  3. IAEA Nuclear Data Section, Live Chart of Nuclides, Cs-137 and Co-60 ground states and decay radiation.
  4. Generated report: docs/accuracy-spec.md; tests: src/validation and decay.test.ts (air-kerma rate constants).