Radiation Lab Field Pro

Guide: from source settings to a reported result

Field Pro · updated 28 October 2026

The workbench takes a reference field from its source settings to a number you can put in a calibration record. This page follows the screen in the order you fill it.

How the screen is laid out

On a wide screen the inputs sit on the left (panels 01–04) and the results on the right (the spectrum chart and panels A–D). Below 980 px the two columns become an Input tab and a Results tab, with the calculate bar fixed to the bottom edge. Every panel title is also an address: #source, #geometry, #readout and #repeatability for the inputs, #spectrum, #characteristics, #operational, #iso and #uncertainty for the results. Such a link opens the right tab at that panel, and Back steps through the panels you visited.

Filling in the four input panels

01 Source

Pick X-ray tube, Cs-137, Co-60 or Import. For a tube, the ISO beam quality list holds all 50 qualities of ISO 4037-1; choosing one fills in the tube voltage, the filter stack and a 2500 mm air path. Edit any of them and the list shows Custom (edited), so a modified beam never carries a standard name. Tube voltage runs from 10 to 400 kV; filter layers can be Be, Al, Cu, Sn, Ta, W or Pb, and the first row is tagged Inherent. For a nuclide, enter the certificate activity in MBq or GBq, the certificate's reference date, the date of the measurement and the distance in metres. Import takes a spectrum as two columns, energy and fluence, separated by commas, tabs or spaces.

02 Geometry and evaluation

The angle of incidence α (0–75°) applies to Hp(10) and H′(0.07); H*(10) is isotropic and ignores it. Leave HVL material on Auto: the half-value layer is then given in the material of the matched ISO quality. The ISO comparison target is auto-matched by default, and the tolerance band starts at ±5%.

03 Readout correction

Enter one chamber reading with the temperature, pressure and relative humidity at the time. The workbench returns kρ, the factor that refers a vented chamber to 20 °C, 101.325 kPa and 65% relative humidity.

04 Repeatability uncertainty

Paste repeat readings separated by spaces, commas or line breaks. With Apply kρ ticked, each reading is multiplied by the panel 03 factor first. Type B components below them are relative standard uncertainties in percent; the first is a 1.0% entry for the reference chamber's certificate.

Calculate or auto-calculate

The workbench starts in manual mode. When an input changes, the button turns into Calculate — inputs changed and the result panels keep showing the previous calculation under a notice, so you never read a half-updated set of numbers. After you press it, each readout shows the change from the previous calculation with ▲ or ▼. Switch on Auto-calculate to update while typing; that choice and the inputs themselves are kept in this browser for your next visit.

Reading the results

Saving, loading and printing

Save session downloads a JSON file holding every input; Load session restores it on any computer. Print report lays out the input conditions and all result panels on paper with the print date. Reset returns every field to the N-80 defaults after a confirmation. Two address parameters open the workbench in a given state: ?preset=N-80 (any quality name) and ?source=Cs-137.

Mistakes that change the answer

The N-80 example runs a complete X-ray field from these panels, and Methods explains what each number is computed from.

Sources

  1. ISO 4037-1:2019, Part 1: Radiation characteristics and production methods — quality list, HVL material of H-60, HVL1 of N-80 and H-10 at 1.0 m and 2.5 m.
  2. JCGM 100:2008, Guide to the expression of uncertainty in measurement, G.3 and G.4 (coverage factor and effective degrees of freedom).
  3. IAEA Nuclear Data Section, Live Chart of Nuclides: Cs-137 half-life 30.08 a.
  4. Screen behaviour as implemented in the workbench source.

Open the workbench →